Sartolab® P20 Pressure Filter 18075----------D, 0.2 µm Polyethersulfone
Sartolab® P20 with 0.2 µm polyethersulfone (PES) is designed for sterile filtration of cell culture media and other solutions. Type 18075 has a Luer-Lock inlet and hose barb outlet. The filters are individually packed and sterilized by ethylene oxide (EO).
- Price (USD):
- $126.00
- Item No.:
- 18075----------D
- Availability:
- pcs
- Pieces
- 10
*Custom/bulk order quotes are provided within 24 hours of request.
Choose your Sartolab® P20 Pressure Filters
Sartolab® P20 is a ready-to-use pressure filter unit for the sterile filtration of 100 mL to 5 L media and aqueous solutions.
Main Benefits:
- Highest flow rates with a large surface of filtration (20 cm²)
- No loss of protein with a low protein affinity membrane
- Low dead volume due to an optimized membrane support
Applications
-
ApplicationClarification
-
Application AreaSterile filtration
Components
-
Filling BellNo
-
PrefilterNo
Connections (Physical)
-
Connector InletFemale Luer Lock
-
Connector OutletHose Barb 6–12 mm Stepped
Dimensions
-
Filter Diameter61 mm
-
Filtration Area20 cm²
-
Housing Dimensions61 x 50 mm
Features
-
Venting FilterNo
General Specifications
-
Flow Rate≥ 400 mL/min at 1 bar
-
Total ThroughputUp to 5 L
-
VentNo
-
pH Range1.0 – 10.0
Materials of Construction
-
Material PrefiltersN/A
-
Membrane MaterialPolyethersulfone (PES)
Operating Conditions
-
Maximum Allowable Working Temperature50 °C (122 °F)
-
Operating Pressure4 bar
Physicochemical Information
-
Bubble Point≥ 3.2 bar (46 psi) with water
-
Burst Pressure≥ 0.5 bar (72.5 psi)
-
Non-Fiber ReleasingThis filter product complies with the title 21 of the Code of Federal Regulations (CFR), section 210.3(b)(6) and 211.72.
-
Pore Size0.2 µm
-
WettabilityHydrophilic
Product Information
-
BrandSartolab® P20
-
Device TypePressure filters without prefilter
-
PackagingSingle pouch packed
-
Sample Type
-
Aqueous
- Solvents according to chemical compatibility
-
Aqueous
Sterility/Purity
-
Delivery ConditionSterile
-
Sterilization MethodEthylene oxide
Volume
-
Hold-up Volume1 mL before bubble point
-
VolumeUp to 5 L