Sartovent DG Cartridge 0.2 µm Size 6
The permanent hydrophobic 0,2 µm PTFE membrane used for Sartovent DG filter cartridges guarantees highest safety and filtration performance for sterile venting of transport containers or similar applications. The filter resist multiple sterilizing cycles and are integrity testable.
- Price (USD):
- $935.00
- Item No.:
- 5181807T6------D
*Custom/bulk order quotes are provided within 72 hours of request.
Choose your Sartovent
Sartovent® DG filter cartirdges are the ideal solution for sterile ventilation of transport containers as well as sterile filtration of fermentation air. Moreover, Sartovent® DG is convenient for sterile filtration of pressurised air and gases which are needed for filling and emptying tanks, containers, filling machines, transport containers, autoclaves and lyophilisators. High flow rates: In spite of its relatively small filtration area of 0.05m², the minimal resistance at filtration of pressurised air enables draining speed of 1000l in 10 min. Permanent hydrophobicity: The PTFE membrane is inherently water-repellent and the most hydrophobic membrane of all used air filter membranes. Even repeated autoclaving or steaming does not affect the membrane. Therefore, sterile filtration of dry or liquid air or different gases is being provided. Pressure-resistant and testable: The cartridges resist high differential pressures in direction of filtration and backwards. The filter structure of the membrane stays attached even at pressure impulses and high flow speeds and provides function testing of the filtration system via integrity testing. Flushing: For the filtration of gases, no flushing is required before use. For the filtration of liquids, flushing could be reasonable for ensuring the absence of any particles in the entire system. Sterilisation: Autoclaving or in-line-steaming up to 134°C is possible whereat the differential pressure should not exceed 0.5 bar. The filter cartridge can be steam sterilized in any direction. A secure hold of the cartridge adapter within the filter housing is the condition for backward steaming.
- Ideal solution for sterile ventilation of transport containers as well as sterile filtration of fermentation air
- Ideal for sterile filtration of pressurised air and gases which are needed for filling and emptying tanks, containers, filling machines, transport containers, autoclaves and lyophilisators
- Made with a hydrophobic PTFE membrane
High flow rates:
In spite of its relatively small filtration area of 0.05m², the minimal resistance at filtration of pressurised air enables draining speed of 1000l in 10 min.Permanent hydrophobicity:
The PTFE membrane is inherently water-repellent and the most hydrophobic membrane of all used air filter membranes. Even repeated autoclaving or steaming does not affect the membrane. Therefore, sterile filtration of dry or liquid air or different gases is being provided.Pressure-resistant and testable:
The cartridges resist high differential pressures in direction of filtration and backwards. The filter structure of the membrane stays attached even at pressure impulses and high flow speeds and provides function testing of the filtration system via integrity testing.Flushing:
For the filtration of gases, no flushing is required before use. For the filtration of liquids, flushing could be reasonable for ensuring the absence of any particles in the entire system.Sterilisation:
Autoclaving or in-line-steaming up to 134°C is possible whereat the differential pressure should not exceed 0.5 bar. The filter cartridge can be steam sterilized in any direction. A secure hold of the cartridge adapter within the filter housing is the condition for backward steaming.
Applications
-
ApplicationAir/Gas Filtration
-
Application Area
-
Bioreactors (Inlet | Outlet)
-
Storage tank venting
-
Filling equipment venting
- Venting
-
Bioreactors (Inlet | Outlet)
-
IndustryPharma / Biotech
Compliance Information
-
Bacterial Endotoxins< 0.18 EU/mL as determined by the LAL test
-
Biological ReactivityAll materials of this filter element meet the requirements of the current USP Biological Reactivity tests <88> for plastics Class VI (Systemic Injection, Intracutaneous and Implantation tests).
-
ExtractablesThe total amount of extractables is well below the limits established by the current USP unter "Sterile Water for Injection".
Dimensions
-
Filter Sizesize 6 (0.05 m² / 0.54 ft²)
-
Filtration Area10"(0.05 m² / 0.54 ft²)
General Specifications
-
Capsule/Cartridge FormatCartridges
-
Filter FormatCartridges
-
Maximum Allowable Back Pressure
-
3.0 bar at 21°C
-
1.5 bar at 121°C
- 0.5 bar at 134°C
-
3.0 bar at 21°C
-
Maximum Diffusion2 mL/min at 0.7 bar l 10 psi at 20 °C
-
Minimum Required Bubble Point1.0 bar | 14.5 psi
Materials of Construction
-
Filter MaterialPolytetrafluorethylene (PTFE)
-
Material o-ring | sealSilicone (optional EPDM or Fluoroelastomer)
-
Membrane MaterialPolytetrafluorethylene (PTFE)
-
Structural CompoundsPolypropylene (PP)
-
Support FleecePolypropylene (PP)
Operating Conditions
-
Maximum Differential Pressure20 °C: 5 bar, 121 °C: 1.5 bar, 134 °C: 0.5 bar
Physicochemical Information
-
Gamma StableNo
-
Pore Size0.2 µm
-
Pore Size Final-filter0.2 µm
-
Retention Rate0.2 µm
Product Information
-
BrandSartovent
-
Pack Size10
-
TypeAir filter
Sterility/Purity
-
Delivery ConditionNon sterile - Autoclavable | Steamable
-
Sterilization MethodAutoclavable & In-Line Steam Sterilizable
Usage Protocols
-
Sterilization Procedures
-
In-Line Steam Sterilization: at 134°C and 2 bar l 29 psi, 20 min. at max differential pressure of 0.5 bar | 7 psi
- Autoclaving: at 134°C, 2 bar | 29 psi, 20 min. or other parameters to be specified
-
In-Line Steam Sterilization: at 134°C and 2 bar l 29 psi, 20 min. at max differential pressure of 0.5 bar | 7 psi