Sartobran® P Cartridge 0.2µm 10 inch
Sartobran® PSartobran® P with its unique low-binding cellulose acetate (CA) membrane is mainly used for the sterile filtration of blood & plasma products and vaccines. The ultra-low binding of the membrane material leads to highest product yield and exceptional filtration performance. Especially large, shear-stress sensitive proteins do not lose their structure and function when filtered through this special membrane. This will improve your product yield even further.
- Item No.:
- 5232807H1----IVG
- Pieces
- 25
*Custom/bulk order quotes are provided within 72 hours of request.
Choose your Sartobran® P
Sterilizing-Grade, 0.2 µm Rated Filters
Sartobran® P Pore size: 0.45 | 0.2 µm (heterogeneous PES double layer)
Bioburden Reduction, 0.45 µm Rated Filters
Sartobran® P Pore size: 0.8 | 0.45 µm (heterogeneous PES double layer)
Your Benefits:
- High product yield due to ultra-low unspecific adsorption
- Excellent filtration results especially for large, shear-stress sensitive proteins
- Exceptional total throughput especially for blood and plasma products
For highest process flexibility all Sartobran® P filters are available for conventional or single use applications. All filter elements are available in various sizes and adapter combinations. Please review the selection guide in the supporting resources literature.
Accessories Information
-
Adapter
-
28 (S-adapter top, locking 3 x bayonet adapter with double O-ring bottom)
-
28 (S-adapter top, locking 3 x bayonet adapter with double O-ring bottom)
-
28 (S-adapter top, locking 3 x bayonet adapter with double O-ring bottom)
-
28 (S-adapter top, locking 3 x bayonet adapter with double O-ring bottom)
-
28 (S-adapter top, locking 3 x bayonet adapter with double O-ring bottom)
-
28 (S-adapter top, locking 3 x bayonet adapter with double O-ring bottom)
-
28 (S-adapter top, locking 3 x bayonet adapter with double O-ring bottom)
-
28 (S-adapter top, locking 3 x bayonet adapter with double O-ring bottom)
- 28 (S-adapter top, locking 3 x bayonet adapter with double O-ring bottom)
-
28 (S-adapter top, locking 3 x bayonet adapter with double O-ring bottom)
Applications
-
Industry8799054758496
Biological Information
-
Bacterial Retention8798947508832
Compliance Information
-
Bacterial Endotoxins8798947344992
-
Biological Reactivity8803105505888
-
Extractables8798986044000
Dimensions
-
Filtration Area0.6 m²
General Specifications
-
Maximum Diffusion15 mL/min at 2.5 bar | 36 psi at 20 °C per 10 inch element
-
Minimum Required Bubble Point2.0 bar | 29 psi
Materials of Construction
-
Material o-ring | seal8799065834080
-
Membrane Material8799042863712
-
Support Fleece8799043191392
Operating Conditions
-
Maximum Differential Pressure8799040963168
Physicochemical Information
-
Non-Fiber Releasing8799064457824
Product Information
-
Brand8798951408224
Usage Protocols
-
Sterilization ProceduresIn-line steam sterilization: 134 °C, 20 min. at max. differential pressure of 0.5 bar.