• Shop
  • Sartopore® 2 | 0.2 µm | Sartoscale

Sartopore® 2 Sartoscale 47 0.2µm | 3pc.

  • Effective Filtration Area: 17.3 cm²
  • Pore Size: 0.45 µm prefilter | 0.2 µm final filter
  • Connectors: Inlet: Sanitary ¾“| Outlet: Sanitary ¾“
  • Membrane Material: Polyethersulfone
  • Pack Size: 3 pieces

Price (USD):
$92.50
Item No.: 
5445307HS--FF--M
Availability: 
Pieces
3

*Custom/bulk order quotes are provided within 72 hours of request.

  • Developed for the execution of reliable filterability samples with 47 mm stamping blades
  • Made from the same non-woven fabric material as the original filter cartridges
  • Ready-to-use single-use test filter unit

Applications

SartoScale units are ideally suited for all types of filterability studies with the aim of selecting the optimal prefilter for an optimal application. Moreover, they can determine the ideal combination of pre- and final filters with the aid of a minimum product volume.

Original filter material

SartoScale units contain the original filter active material of the corresponding filter cartridges to ensure the repeatability of test results.

Ready to Use

Using SartoScale during filtration studies saves time which would otherwise be used for the installation preparations for the punch piece in the stainless steel filter brackets and prevents possible installation errors.

Scalability

After the material selection or the definition of a pre/final filter combination with the SartoScale units, a scale-up method should be carried out for the flow rate and total flow using pleated capsule units of small scales (e.g. capsule size 4).

Optimized design

SartoScale units have an extremely low residual volume for performing filterability samples with minimized product volumes.

Process Reliability

SartoScale test filter units are equipped with suitable filter membranes for integrity testing so that their integrity can be determined using the bubble point test in order to ensure absolute reliability of the test results.

Zero-T test system

We recommend using the SartoScale units together with the Zero-T Filter Test System to make the filtration studies as effective as possible. The Zero-T system consists of hardware and software modules which make it easy to handle and install Sartoscale test filter units. Automatic data collection is achieved by connecting a frame to a laptop. The software automatically analyzes the incoming data for the scale-up calculations.

Compliance Information

  • Bacterial Endotoxins
    < 0.18 EU/mL as determined by the LAL test
  • Biological Reactivity
    All materials of this filter element meet the requirements of the current USP Biological Reactivity tests <88> for plastics Class VI (Systemic Injection, Intracutaneous and Implantation tests).
  • Extractables
    The total amount of extractables is well below the limits established by the current USP unter "Sterile Water for Injection".

Connections (Physical)

  • Connector(s)
    Sanitary 0.75" | Sanitary 0.75"

Dimensions

  • Filter Size
    17.3 cm²
  • Filtration Area
    17.3 cm²

General Specifications

  • Capsule/Cartridge Format
    Sartoscale 47
  • Minimum Required Bubble Point
    3.2 bar | 46 psi

Materials of Construction

  • Membrane Material
    Polyethersulfone (PES)

Operating Conditions

  • Maximum Differential Pressure
    20 °C: 5 bar, 80 °C: 2 bar

Physicochemical Information

  • Non-Fiber Releasing
    This filter product complies with the title 21 of the Code of Federal Regulations (CFR), section 210.3(b)(6) and 211.72.
  • Pore Size Final-filter
    0.2µm
  • Pore Size Pre-filter
    0.45µm

Product Information

  • Brand
    Sartopore® 2
  • Pack Size
    3
  • Type
    Sartoscale 47 | Connectors: Sanitary | 3 pc.

Usage Protocols

  • Sterilization Procedures
    Autoclaving: 1 cycle at max. 134 °C, 2 bar | 29 psi, 30 min.