Sartopore® 2 Gamma Midicaps® 0.2µm Size 9 | 4pc.
- Effective Filtration Area: 0.2 m² | 2.2 ft²
- Pore Size: 0.45 µm prefilter | 0.2 µm final filter
- Connectors: Inlet: Sanitary 1 ½“ | Outlet: Sanitary 1 ½“
- Sterilization: Gamma-irradiatable
- Membrane Material: Polyethersulfone
- Pack Size: 4 pieces
________________________________________
Connector explanation:
O = Hose Barb ½“ | S = Sanitary 1 ½“ | H = Hose Barb ¼“ | F = Sanitary ¾“
Size explanation:
Size 7 = 0.065 m² | Size 8 = 0.13 m² | Size 9 = 0.26 m² | Size 0 = 0.52 m²
- Price (USD):
- $1,410.00
- Item No.:
- 5445307H9G-SS--A
- Pieces
- 4
*Custom/bulk order quotes are provided within 72 hours of request.
Choose your Sartopore® 2
Sartopore® 2 Gamma Midicaps® were developed for use with bag containers prior to sterilization using gamma irradiation. Applications
Typical applications include the sterile filtration and mycoplasma disposal of:
- Biologicals
- Pharma
- Cell culture media (with and without serum)
- Serum
- Buffer
- Culture media components
Compatibility
Sartopore® 2 Gamma Midicaps® have been developed for sterilization using gamma irradiation < in 50 kGy irradiation doses. The Sartopore® 2 polyethersulfone membrane has a high chemical compatibility between pH 1 to 14, making Sartopore® 2 Gamma Midicaps® ideally suited for the filtration of high and low pH buffers in pharmaceutical | Biotech Industry
Performance
Through "built-in prefiltration" in the form of a heterogeneous two-layer membrane, Sartopore® 2 Gamma Midicaps® achieve outstanding total flow and excellent flow rates.
Flexibility
Sartopore® 2 0.2 μm Midicaps® are supplied with different filtration areas of 500 cm² | 0.5 ft² to 0.45 m² | 5 ft² for easy adaptation to every filtration process regardless of batch size.
Microbiological Retention
Sartopore® 2 Gamma Midicaps® are completely validated as a sterile filter in accordance with HIMA ASTM F-838-05 guidelines. In addition, for mycoplasma retention, Sartopore® 2 Gamma Midicaps® are validated with an LRV of 7 for acholeplasma Laidlawii.
Quality
Each individual item was assigned before publication by a Bubblepoint and a diffusion test, integrity tested to ensure absolute reliability.
Documentation
Sartopore® 2 Gamma Midicaps® have been designed, developed and manufactured to comply with an ISO 9001 certified quality management system. A validation manual is available that meets regulatory requirements.
Applications
-
ApplicationHigh capacity platform for conventional sterile filtration
-
IndustryPharma / Biotech
Biological Information
-
Bacterial RetentionQuantitative Retention of 1 × E7 CFU/cm² Brevundimonas Diminuta ATCC® 19146 per ASTM methodology
Compliance Information
-
Bacterial Endotoxins< 0.25 EU/mL as determined by the LAL test
-
Biological ReactivityAll materials of this filter element meet the requirements of the current USP Biological Reactivity tests <88> for plastics Class VI (Systemic Injection, Intracutaneous and Implantation tests).
-
ExtractablesThe total amount of extractables is well below the limits established by the current USP unter "Sterile Water for Injection".
Connections (Physical)
-
Connector Inlet1.5" Tri-clamp 50 mm
-
Connector Outlet1.5" Tri-clamp 50 mm
-
Connector(s)Sanitary 1.5" | Sanitary 1.5"
Dimensions
-
Filter Sizesize 9 (0.2 m² / 2.2 ft²)
-
Filtration Area0.2 m²
-
Outer Diameter108.5 mm (77 mm without Connection / Valves)
General Specifications
-
Capsule/Cartridge FormatMidicaps®
-
Maximum Allowable Back Pressure2 bar (29 psi) at 20°C; 1.5 bar (21.8 psi) at 50°C
-
Maximum Diffusionat 20 °C and 2.5 bar | 36 psi: 0.05 m² | 0.5 ft²: 4 mL/min ; 0.1 m² | 1 ft²: 5 mL/min; 0.2 m²| 2 ft²: 7 mL/min; 0.45 m² | 4.5 ft²: 14 mL/min
-
Minimum Required Bubble Point3.2 bar | 46 psi
-
Vent ValveStandard
Materials of Construction
-
Material o-ring | sealSilicone (SI)
-
Support FleecePolyester (PET)
Operating Conditions
-
Maximum Differential Pressure20 °C: 5 bar, 50 °C: 3 bar
-
Maximum Operating Pressure5 bar (72.5 psi)
Physicochemical Information
-
Non-Fiber ReleasingThis filter product complies with the title 21 of the Code of Federal Regulations (CFR), section 210.3(b)(6) and 211.72.
-
Pore Size Final-filter0.2 µm
-
Pore Size Pre-filter0.45 µm
Product Information
-
BrandSartopore® 2
-
Pack Size4
-
TypeSize 9 | Connectors: SS
Sterility/Purity
-
Sterilization MethodGamma Irradiatable
Usage Protocols
-
Sterilization ProceduresGamma Irradiation: with 50 kGy, 1 Cycle