CIMac™ EDA 0.1 mL Analytical Column (1.3 µm)

CIMac™ Analytical Columns combine all elaborate characteristics of CIM monoliths at analytical scale and are your perfect tool for analysis and control of manufacturing processes, ideal for process analytical technology (PAT).

Designed for fast and reproducible HPLC monitoring and quantitation of proteins, viruses, pDNA, and other large biomolecules, they allow you in a matter of minutes to achieve
rapid high-resolution
flow-independent separations.

Important: Only available in conjunction with CIMac™ Analytical Software

Item No.: 
110.5116-1.3
Piece
1

*Custom/bulk order quotes are provided within 72 hours of request.

New generation of CIMac™ Analytical Columns

Working principle

The CIMac® EDA 0.1 ml Analytical Columns are weak anion-exchange monolithic columns, that bind biomolecules that are predominantly negatively charged over the pH working range of 3-9.

A unique monolith

In contrast to DEAE or QA monoliths it contains free primary amino groups, resulting in a promotion of not purely ion-exchange interaction.

EDA (ethylenediamine) features:

  • Weak anion exchanger
  • Working range, pH 3–9
  • Binds molecules with predominantly negative charge, repels molecules a predominantly positive charge
  • Use for capture, intermediate, or polishing steps in either bind-elute or flow-through mode

Key benefits

  • CIMac™ has high capacity for plasmid DNA
  • Sanitizable with 1 M NaOH
  • Saves CO2

Please also check our new CIMac™ Analytical Software

Applications

  • Application
    Analytical Columns

Connections (Physical)

  • Connector(s)
    10-32 UNF coned port, 0.0625" OD tubing connection

Dimensions

  • Channel Size
    1.3 µm
  • Diameter
    5.2 mm
  • Length
    4.95 mm

Features

  • Reusable
    Yes

General Specifications

  • Flow Rate
    0.2 – 3 mL/min | 1 – 15 cm/min | 2 – 30 CV/min

Operating Conditions

  • Maximum Operating Pressure
    15 MPa | 150 bar | 2175 psi
  • Recommended Operating Temperature
    4 °C (39 °F) to 40 °C (104 °F)
  • pH Stability
    Working range 2–10, cleaning in place 1–14

Physicochemical Information

  • Bed Volume
    0.1 mL
  • Ligand
    EDA (weak anion exchanger, ethylene diamine)

Product Information

  • Brand
    CIMac™